Kinematic PET: Turning Total-Body List-Mode Data Into Video

Supplementary Fig. 11 of arXiv:2611.11052 by Reader et al.: maximum intensity projections of one 33 ms total-body FDG PET frame, noisy MLEM with spatial smoothing on the left and HEM+VAST on the right showing vessels, heart and kidneys.

Total-body PET scanners collect so many counts that the list-mode file of a routine dynamic scan holds far more timing detail than anyone uses. Conventional dynamic PET still bins that data into frames of seconds to minutes, because cutting it finer gives images made mostly of noise. Andrew J. Reader and colleagues at King’s College London, the Guy’s and St Thomas’ PET Centre and UC Davis go the other way in arXiv:2611.11052 (v1 8 November 2026, revised 24 November 2026). They reconstruct a human FDG scan from a Siemens Biograph Vision Quadra in 33 ms frames, about 30 volumes per second, and play the result back as a video of the whole body.

They call the task kinematic PET. The goal is to watch things move: the tracer bolus travelling through the vessels, cardiac contraction and respiratory motion of the spleen and kidneys, all without gating. The method behind it is a two-stage loop named HEM+VAST.

What goes in and what comes out

The input is time-of-flight list-mode data. In the human demonstration that is 30 s of data recorded 28 to 58 s after injecting 214.51 MBq of [18F]FDG, split into 910 frames of 33 ms. Each frame holds only about 300k to 500k prompt coincidences, against several million voxels per volume. The output is 910 full 3D volumes on a 200 x 200 x 350 grid with 3 mm voxels, covering a 600 x 600 x 1050 mm field of view.

The authors built this output for looking at motion and early tracer transport and say plainly that it is not quantitative. Attenuation, scatter, randoms, normalisation and dead-time corrections were all left out of this reconstruction. Read the frames as a cine for visual assessment. SUV measurements are out of scope for this version.

How HEM+VAST works

The reconstruction starts with six independent list-mode EM iterations per frame, which gives a sequence of very sparse, speckled volumes. Then it alternates two stages.

Hierarchical EM (HEM). Ordinary EM computes a multiplicative correction image for each frame from that frame’s own events. At 33 ms those corrections are dominated by noise. HEM pools the corrections over overlapping temporal blocks at four levels (the full scan duration, 4 s blocks, 2 s blocks and the single frame) and blends them, with the frame level weighted at 40% and each other level at 20%. Each frame still gets its own image, but the update borrows counts from its neighbours in time. Voxels that receive no events in a frame get a neutral update instead of being pushed to zero, and a global relaxation (20% data consistency in the main setting) plus clipping keeps the updates gentle.

Validation-adaptive space-time regularisation (VAST). For every 11 x 11 x 11 voxel patch in a frame, VAST draws 32 random source patches from frames up to 9 frames away (skipping the immediate neighbours) and up to 8 mm away in space. Each source is mixed with the target patch, and the candidates are scored against a validation patch built by averaging the frame just before and the frame just after. The unchanged patch competes too, so VAST can decline to change anything. Overlapping winners are averaged. Because the validation frames are excluded from the sources, the choice is checked against data that did not produce the candidate.

Repeating HEM then VAST lets useful information spread step by step through space and time. An optional post-processing step, temporal block sum correction, rescales each voxel’s time course so that the sum over all frames matches the sum of the initial EM images. This corrects the spatial blurring that builds up over the cycles and leaves the shape and peak timing of each time course unchanged.

What the comparison figure shows

The featured image uses Supplementary Fig. 11 of the paper. Both panels are maximum intensity projections of the same 33 ms frame. On the left is MLEM after 6 iterations with 10 mm spatial smoothing, on the right 12 cycles of HEM+VAST, both with block sum correction. The MLEM frame is covered in hot speckle, and the kidneys and heart sit in a grainy haze. The HEM+VAST frame shows a bright arm vein, the great vessels, both kidneys and the body outline against a clean background. The paper’s Fig. 7 makes the same comparison on a transverse slice through the heart, and the authors posted videos of the full reconstruction and of the side-by-side comparison.

What the numbers say

Most of the measurement happens on a digital phantom with moving boluses at 35 to 150 mm/s, a beating heart at 90 beats per minute and a breathing lung at 15 breaths per minute, about 170k true coincidences per 33 ms frame. At a matched background noise of 10% spatial coefficient of variation, the fastest bolus came out with a temporal width of 151 ms, against 86 ms in the ground truth. Converted to a temporal resolution estimate, that is 124 ms for HEM+VAST against 171 ms for VAST alone.

On the patient data, a region over the heart shows 13 oscillations between 35 and 45 s, which the authors read as a heart rate of about 78 beats per minute, and a neck vessel shows fluctuations consistent with breathing.

Why an imaging platform team would care

For a viewer, a kinematic series is a new kind of object: hundreds of full-body volumes that only make sense played as a cine, usually as a rotating MIP. The human run here is 910 frames of 14 million voxels each, about 12.7 billion voxels for 30 s of scan, so streaming, caching and MIP rendering on the fly decide whether anyone can scrub through it. A useful viewer would show the cine next to the summed static image and let the user drop a region and see its time-activity curve, which is how the paper reads cardiac and respiratory motion in Figs. 8 and 9.

The authors also point at two downstream uses. Motion seen directly in the images could feed motion estimation for conventional quantitative reconstruction, and the bolus passage could support image-derived physiological measures of organ motion and tracer transport. The paper proposes both but does not test them.

Limits to keep in mind

  • The frames are not quantitative. No attenuation, scatter, randoms, normalisation or dead-time modelling was applied, and HEM’s relaxation and clipping move it away from a pure maximum-likelihood method.
  • The human demonstration is a single anonymised FDG acquisition. All the quantitative evaluation is on a simulated phantom.
  • HEM+VAST has not yet been compared with kernel EM, HYPR4D, deep image prior or other 4D PET methods at the same 33 ms frames and matched noise. The authors name this as the next step.
  • The cardiac oscillation in the heart region time course is smaller for HEM+VAST than for the EM baselines. The authors note that the noise levels were not matched, so the comparison is not settled.
  • Compute is heavy. On an RTX 5090 with 128 GB of RAM, one cycle for 910 frames took about 3.5 hours (up to 90 min for HEM, 50 to 60 min per VAST sweep, with two sweeps per cycle), and the figures and videos use 12 to 24 cycles.
  • There are many settings (block levels and weights, relaxation, patch size, search range, trials), tuned on the phantom.

Code, weights and licence

When we checked on 25 November 2026, the paper did not link any code, and we found no public implementation on GitHub. HEM+VAST is an iterative reconstruction algorithm with no trained network, so there are no weights to release. The patient data come from the ethically approved TOTEM study at the King’s College London and Guy’s and St Thomas’ PET Centre and are not public. The paper is released under a Creative Commons Attribution 4.0 (CC BY 4.0) licence on arXiv, which is why we can reproduce its figure here with credit.

Source

Andrew J. Reader, Gary J. R. Cook, Paul K. Marsden, Sarah-May Gould, Alexander Hammers and Simon R. Cherry, “Kinematic Total-Body 3D PET Video Reconstruction with 33-ms Frames via Hierarchical List-Mode EM and Validation-Adaptive Space-Time Regularisation”, arXiv:2611.11052 (v2 24 November 2026), preprint, CC BY 4.0. Affiliations: School of Biomedical Engineering and Imaging Sciences, King’s College London; Guy’s and St Thomas’ PET Centre, London; Department of Biomedical Engineering, University of California, Davis. Featured image: two maximum intensity projection panels of one 33 ms frame (MLEM with spatial smoothing and HEM+VAST) cropped from Supplementary Fig. 11 of the paper, with labels added.