CANON-QSM: Iron and Myelin Source Maps in Under Two Minutes

Fig. 6 of arXiv:2611.26561 by Contreras-Zúñiga, Tejos and Milovic: sagittal composite susceptibility and paramagnetic component maps of a healthy volunteer at 3 T, reference DECOMPOSE-QSM on the left and CANON-QSM on the right, with the 0 to 0.1 ppm scale.

Quantitative susceptibility mapping (QSM) turns the phase of a gradient-echo MRI scan into a map of tissue magnetic susceptibility, and research groups use it as a marker of brain iron and myelin in aging and neurodegenerative disease. A standard QSM map gives one number per voxel. Iron is paramagnetic and myelin is diamagnetic, so when both sit in the same voxel their effects partly cancel, and a voxel rich in both can look the same as a voxel with neither. Ignacio Contreras-Zúñiga, Cristian Tejos and Carlos Milovic at the Biomedical Imaging Center of the Pontificia Universidad Católica de Chile (arXiv:2611.26561, posted 19 November 2026) present CANON-QSM, a solver that splits that single number into separate paramagnetic and diamagnetic maps for a whole brain in under two minutes.

CANON-QSM keeps the physics model of an existing method, DECOMPOSE-QSM, and solves it in a form a GPU can run quickly and that avoids fitting failures in large veins.

What source separation needs from the scan

The input is a multi-echo gradient-echo (GRE) acquisition, with magnitude and phase saved for every echo. No extra spin-echo sequence is needed, unlike χ-separation, which combines QSM and R2* with an R2 or R2′ map. The model has five unknowns per voxel and each echo gives two real measurements, so at least three echoes are required. The in vivo protocol in the paper used five echoes at 3 T, a first echo time of 7.2 ms with 6.2 ms spacing, and 0.60 x 0.60 x 1.0 mm voxels.

Before the solver runs, each echo is turned into its own QSM map. The paper’s pipeline uses ROMEO for phase unwrapping, PDF for background field removal, WH-QSM for harmonic remnants and DEMIX-QSM for the per-echo dipole inversion. The authors expand the signal in powers of echo time and show why the maps must be per echo. Standard QSM and a mono-exponential R2* fit capture only first-order quantities, the net susceptibility and the total decay. The two sources are separated by second-order effects. In a mixed voxel the apparent susceptibility drifts as echo time grows, and the log-magnitude decay curves. An echo-combined QSM map averages that drift away.

How the solver works

Each voxel is modeled as three pools of water: one near paramagnetic sources, one near diamagnetic sources and one neutral. Each pool has a concentration, the two source pools have a susceptibility, and all three share a baseline relaxation rate. DECOMPOSE-QSM fits this by alternating over three blocks of unknowns and solving each block with a general-purpose constrained MATLAB solver. CANON-QSM keeps the same alternation but replaces the inner solvers.

Concentrations in closed form. With the susceptibilities and relaxation fixed, the measured signal has to be the closest point inside a triangle whose corners are the three pool signals. The authors show the answer is always one of four candidate points, so this block needs no iterations. A small floor on every concentration keeps any pool from vanishing early in the fit, and it shrinks over the iterations so it does not bias the final answer.

Relaxation in closed form. The shared relaxation rate only scales signal magnitude, so it reduces to a log-linear fit through the origin.

Susceptibilities by Newton-type steps. The remaining block has two unknowns. The residual compares phases after wrapping the difference, so a 2π jump at a late echo or in a large vein does not add a spurious penalty. Each step is a 2 x 2 Levenberg-Marquardt system solved directly. Because this block is not convex, every voxel runs 20 starts in parallel, one deterministic and the rest random, and keeps the best-scoring one.

Every voxel is independent once the per-echo QSM maps exist, so the whole brain runs as one large batch of tiny problems in PyTorch on a GPU.

What it outputs on a scan

Three maps in ppm: a composite susceptibility map, a paramagnetic component susceptibility (PCS) map and a diamagnetic component susceptibility (DCS) map. In brain tissue the paramagnetic pool is dominated by iron in ferritin and the diamagnetic pool by myelin, which is why these maps are read as iron-weighted and myelin-weighted views. The featured image shows Fig. 6 of the paper, a healthy volunteer at 3 T in sagittal view. On the left is the reference DECOMPOSE-QSM fit, which the authors report produces fitting artifacts in the great vein of Galen in the composite and PCS maps, assigning paramagnetic venous voxels to the diamagnetic pool. On the right is CANON-QSM on the same data, where those venous artifacts are absent.

What the results show

On the 352 x 352 x 160 in vivo volume, CANON-QSM finished in 113 seconds on an RTX 5090 GPU with 1.68 GB of peak memory. The MATLAB reference implementation of DECOMPOSE-QSM took about 5 hours on an 18-thread CPU. On a realistic 7 T brain phantom with known ground truth, CANON-QSM’s composite error at a signal-to-noise ratio of 100 was 6.0% against 13.8% for the reference run at its recommended 10 iterations, and it scored higher on the XSIM similarity measure at every noise level tested. The reference solver was slightly more accurate in venous blood and the dentate nucleus in the noiseless phantom.

Where it falls short

  • One volunteer. The in vivo evidence is a single healthy subject at 3 T. There is no patient data, no histology and no second scanner or protocol. The authors say a larger cohort and histological or clinical data are needed.
  • Same model, same assumptions. CANON-QSM inherits the static dephasing regime, isotropic susceptibility and a fixed count of three pools from DECOMPOSE-QSM. It also inherits any error in the per-echo QSM maps it is fed.
  • A per-dataset bound. The maximum allowed susceptibility is set per dataset for in vivo data, since there is no ground truth to derive it from.
  • Cortex at 3 T needs more restarts. With the default 20 starts, only 79.5% of cortical gray matter voxels in the 3 T phantom reached the reference solution, against at least 99.9% for every tissue at 7 T. Raising the starts to 64 brought every tissue to at least 96%.
  • The speed comparison mixes hardware. The 113 seconds is a GPU run in PyTorch, and the 5 hours is a CPU run in MATLAB, so part of the gap is hardware and implementation rather than the math alone.
  • Short echo trains carry little signal. The paper notes that the second-order information driving the separation fades quickly when the echo train is short.

Where it fits for an imaging platform

For a neuro research platform or a viewer that already handles QSM, CANON-QSM turns whole-brain source separation from a job of several hours into a solver step of about two minutes on one GPU, not counting the per-echo QSM preprocessing. The outputs are ordinary 3D maps in ppm and can be stored as derived series next to the GRE data, with a fixed window such as 0 to 0.1 ppm for the PCS and |DCS| maps as in the paper. Most of the integration work is upstream. The site has to save phase for every echo, and the platform has to run a full per-echo QSM pipeline before the solver sees any data.

These maps are still research measures. Nothing in the paper supports using them for diagnosis.

Code, weights and licence

The paper’s data availability statement points to ignacontreras/CANON-QSM on GitHub. When we checked on 23 November 2026, the repository held only a README and an MIT licence, with the code marked as coming soon and promised to include the solver, the phantoms and the experiment scripts. CANON-QSM is an optimization method, so there are no trained weights to release. The in vivo data are a single volunteer scanned under institutional ethics approval and are not released. The phantom comes from Ridani, De Leener and Alonso-Ortiz (Magnetic Resonance in Medicine, 2026). The paper itself is released under CC BY 4.0.

Source

Ignacio Contreras-Zúñiga, Cristian Tejos and Carlos Milovic, “Susceptibility source separation via Closed-form Alternating Newton-type Optimization”, arXiv:2611.26561 (v1 19 November 2026), preprint. Affiliations: Biomedical Imaging Center and Department of Electrical Engineering, Pontificia Universidad Católica de Chile; Millennium Institute for Intelligent Healthcare Engineering (C. Tejos). Featured image: sagittal composite and PCS panels from Fig. 6 of the paper (in vivo 3 T, DECOMPOSE-QSM vs CANON-QSM) with the figure’s PCS colour scale, cropped from the PDF under CC BY 4.0, with labels added.

We build custom medical imaging platforms — advanced DICOM viewers, AI segmentation, and the clinical systems around them.

Copyright © 2026 PYCAD. All Rights Reserved.