Adaptive radiotherapy runs on three kinds of images of the same patient. The plan is drawn on a CT, the daily setup image is a cone-beam CT (CBCT) from the linac’s on-board imager, and more and more centres add MRI. Each of them needs organ contours, and the contours should agree, because the dose check on the day compares structures on the CBCT with the same structures on the plan. Today’s whole-body segmenters were trained mostly on diagnostic CT. On a noisy CBCT with scatter, truncation and metal streaks they lose organs, and two tools rarely use the same label conventions.
IMPACT-Seg, posted on 24 November 2026 as arXiv:2611.32903, is one network that segments 126 structures on CT, CBCT and MRI with the same weights and no modality flag. It comes from Cédric Hémon, Christopher Kurz and Guillaume Landry (Department of Radiation Oncology, LMU University Hospital, Munich), Valentin Boussot and Matthew McCormick (Fideus Labs, Durham, North Carolina) and Lukas Zimmermann (Medical University of Vienna). The code, the weights and most of the training data are public.
What it hangs on a scan
You give it a volume and it returns one label map on the grid of that volume. The 126 structures run from the head and neck to the pelvis: abdominal and pelvic organs, bowel, the five lung lobes, the four heart chambers with the myocardium and great vessels, every vertebra and rib numbered by level, major bones and trunk muscles, and head-and-neck organs at risk. Each voxel carries a single label, so enclosing classes were dropped. The heart is its chambers plus myocardium, and the larynx is glottis plus supraglottis. Five small outputs (cochleae, optic chiasm, optic nerves) were not learned reliably at the working resolution and are suppressed.
How the training set was built
The networks are standard residual-encoder U-Nets. The new work is in how the training set was assembled.
- One label space. The authors pooled public cohorts into 8,290 volumes: 4,366 CTs from 15 cohorts, 781 MRIs (AMOS22, TotalSegmentator MRI, a whole-body PET/CT/MRI dataset and SynthRAD) and 279 real CBCTs from the SynthRAD 2023 and 2025 challenges, plus simulated CBCTs. All of them were mapped to the label definitions of CADS, an existing whole-body CT model.
- Filling the gaps. Each public dataset labels only a few organs. Missing structures were filled with predictions from existing models (CADS for CT, NV-Segment-CTMR and TotalSegmentator MRI for MRI), refined over several training generations, then reviewed by eye and corrected by hand or with nnInteractive. MRI needed the most fixing: 218 of 374 TotalSegmentator MRI cases were edited.
- Simulated CBCT with free labels. From 955 TotalSegmentator CTs, the Reconstruction Toolkit (RTK) projected cone-beam acquisitions with five scanner profiles taken from the SynthRAD centres (three Elekta XVI configurations, one IBA Proteus, one Varian TrueBeam imager) and reconstructed them back onto the CT grid. Each CT got three versions with random noise, scatter, detector blur, shading, truncation, miscentring, breathing motion and metal artefacts. Because the CBCT sits on the CT grid, it inherits the CT labels voxel for voxel.
- Joint training. A 2.5D network (five axial slices in, the middle one out) and a 3D network (128-voxel cubes) were trained on CT, CBCT and MRI mixed in every batch, at 1.5 by 1.5 by 2 mm, each in five folds. The loss does not ask the modalities to agree; they share only the labels and the weights.
What the evidence shows
The gains show up on CBCT. On 139 real SynthRAD CBCTs, the 3D network reached a mean Dice of 0.842 against 0.671 for the best comparator, NV-Segment-CTMR. On 497 simulated CBCTs whose source CTs the competitors had seen in training, its Dice fell by 0.022 from CT to CBCT, against 0.153 to 0.178 for TotalSegmentator, CADS and NV-Segment-CTMR. The featured image shows the real pelvic CBCT that the authors picked as the median case, with both hip prostheses in the field: 0.85 for IMPACT-Seg 3D, 0.79 for NV-Segment-CTMR, 0.50 for CADS and 0.47 for TotalSegmentator on that slice’s shared structures.
For an external clinical test, the authors used the Memorial Sloan Kettering pancreatic radiotherapy collection on TCIA (planning CT plus two treatment CBCTs for each of 40 patients). All tools scored about the same on the planning CT, but on the daily CBCTs the 3D network reached a gastrointestinal Dice of 0.735 against 0.634 for TotalSegmentator.
For adaptive work, cross-modality agreement counts as much as accuracy. Segmenting the planning CT and the registered real CBCT of the same patient separately, the two IMPACT-Seg masks overlapped at 0.881, against 0.492 for TotalSegmentator. Agreement says nothing about whether both masks are right, so the authors always report it next to accuracy.
End to end on an RTX A6000, a whole-body CT took 8.0 s with the 2.5D model and 14.8 s with the 3D one, and a CBCT or MRI about 2 s, with under 6 GB of GPU memory for one fold. TotalSegmentator took 328 s in the configuration used for the comparison.
Where it fits in a viewer or clinical platform
For a radiotherapy or image-guidance product, one model that gives the same structure names and conventions on the plan CT, the daily CBCT and an MRI removes a lot of glue code: no per-modality model choice, no label remapping between tools. Running on CBCT and MRI in about 2 s is short enough for an online adaptive session. The released app takes a NIfTI file (pip install impact_seg_konfai, then impact-seg-konfai segment fine-3d), so a DICOM pipeline has to convert the series first and turn the label map back into RTSTRUCT or DICOM SEG. There is also a 3D Slicer extension.
The model has two side uses. The disagreement between the five fold models flagged structures with Dice under 0.7 on external CT and MRI with an AUROC of 0.92 to 0.94, and reviewing the 20% most uncertain structures caught 75% (2.5D) and 81% (3D) of failures. In a review screen, that becomes a “check these contours first” list. The network’s internal features also work as a similarity measure for CT to CBCT and MRI to CT registration: the authors’ pipelines built on them ranked first on both multimodal Learn2Reg tasks.
Limits
- On diagnostic CT and MRI it trails the specialists. On 247 external TotalSegmentator v3 CTs its 3D Dice was 0.831 against 0.932 for TotalSegmentator and 0.890 for CADS, with the largest gaps on structures under 10 ml. The authors note that TotalSegmentator scored above 0.99 on 9.7% of structure pairs there, which hints that its own outputs ended up in the references, but they do not claim this explains the gap.
- Vertebra and rib numbering causes much of the CT gap. The 2.5D network lacks cranio-caudal context and can shift lumbar levels; the 3D network identified the correct level 97% of the time on held-out CT against 98% for TotalSegmentator. On a real thoracic CBCT failure case, both networks drew the ribs and vertebrae but shifted them by one level.
- External whole-body MRI is weaker (3D Dice 0.778 against 0.843 for TotalSegmentator MRI), because whole-body scans are rare in its MRI training data.
- On the pancreatic CBCTs, rigidly propagated planning contours did better on the separate small bowel and stomach-duodenum structures, and agreement with the expert contours (0.795) stayed below the agreement between two observers (0.847).
- Many references are pseudo-labels from other models, reviewed but not fully expert-audited. Structures missed entirely leave no uncertainty signal. Correction time, contour acceptability and dose impact have not been measured.
Licence, code and status
The paper is a 60-page arXiv preprint under the standard arXiv licence, not peer reviewed. The training and inference code (built on the KonfAI framework) is on GitHub at HEMONCedric/IMPACTSeg under Apache 2.0, and the 2.5D and 3D weights, five folds each, are on Hugging Face (HCedric/IMPACTSeg), also Apache 2.0. The curated dataset is on Hugging Face as well, but each folder keeps the licence of its source dataset, and several (KiTS, LiTS, CHAOS, HaN-Seg, SynthRAD, TotalSegmentator MRI) are non-commercial. Check those terms before you build a product on the data. The work was funded by the Bavarian Cancer Research Center (BZKF).
Source: Hémon, Boussot, McCormick, Zimmermann, Kurz and Landry, “IMPACT-Seg: Consistent Anatomical Segmentation and Representation Learning Across CT, CBCT, and MRI”, arXiv:2611.32903. Featured image built from the real-CBCT row of Fig. S5 of the preprint with credit to the authors.